Determination of Sulfate concentration in process water

Sulfate analysis allows for quick and easy monitoring of the analyte concentration present in wash water. Sulfate monitoring is critical to preserving process quality and optimizing operational efficiency.

Method

Test type: End Point.

Testing time: 1 minute.

Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab® and 3 tests for CDR GalvanLab® Junior.

13%

Sample treatment

The sample can be used as is.

Principle of the test

Sulfate ions react with the chromogen in the reagent and generate colored complexes whose absorbance is proportional to the concentration of the analyte in the sample.

Calibration curve

Calibration curve was obtained by correlating the CDR method with the reference method.

Sulfate Analysis in Process Water: Calibration Curve of the CDR GalvanLab® instrument with the reference method.

Reagent test Kits

Code *400120: 100 test
Code *400125: 10 test
The reagents have a shelf-life of 12 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Sulfate [10-250] mg/L10 - 250 mg/L1.000 mg/L7 mg/L

The chemical analysis instrument for electroplating baths

CDR GalvanLab®

  • Complete analysis panel, supplied already configured
  • Up to 16 determinations simultaneously
  • Possibility of carrying out analyses of the same sample
  • Integrated printer
  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)

CDR GalvanLab® Jr

  • Partial analysis panel, customisable configuration
  • Up to 3 determinations simultaneously
  • Wireless connection to external printer
  • USB connections