Determination of Gold in Gold Flash bath

The analysis of the gold concentration in Gold Flash plating baths is crucial to optimize the electrodeposition parameters (e.g. time, deposit thickness), ensuring uniform deposits (0.2-0.3 µm) and preventing aesthetic or functional defects in the finished products. Determining the quantity of gold ensures high yield and compliance with environmental regulations. A punctual control on the bath allows to precisely monitor the gold concentration, avoiding losses of the precious metal and reducing waiting times for analysis and corrections.

Method

Test Type: End Point.

Time testing: 1 minute
Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab®.

Sample Treatment

The sample can be used as is.

Test principle

Gold content is determined by its reaction with a chromogen that develops a coloration, the optical intensity of which is proportional to the concentration of Au in the analyzed sample.

Calibration curve

The calibration curve was obtained by correlating the CDR method with the reference method.

Calibration curve of Gold on Gold Flash baths

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Gold on Gold Flash0.1 - 1.2 g/L0.010 g/L0.02 g/L

Reagent test Kits

Code *400210: 100 test
Code *400215: 10 test
The reagents have a shelf-life of 4 months

The chemical analysis instrument for electroplating baths

CDR GalvanLab®

  • Complete analysis panel, supplied already configured
  • Up to 16 determinations simultaneously
  • Possibility of carrying out analyses of the same sample
  • Integrated printer
  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)