Determination of Iron concentration in Pd-Fe plating bath

Iron analysis allows the quantification of Iron concentration in Pd-Fe plating baths. If the amount of Fe in the plating bath is less than the expected amount, the plating process may be suboptimal, possibly leading to problems with aesthetic defects (e.g. glazing) in the finished products. Therefore, monitoring the concentration of Fe in Pd-Fe baths is critical to ensure the good performance of plating processes.

Method

Test type: End Point.

Time testing: immediate reading
Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab®.

Sample treatment

The sample can be used as is.

Test principle

The iron content is determined by its reaction with a chromogen that develops a coloration, the optical intensity of which is proportional to the Fe concentration in the analyzed sample.

Calibration curve

The calibration curve was obtained by correlating the CDR method with the reference method.

Calibration curve Iron in Pd-Fe baths

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Iron on PdFe bath [0.05-0.3] g/L0.05 - 0.3 g/L0.001 g/L0.003 g/L

Reagent test Kits

Code *400020: 100 test
Code *400025: 10 test
The reagents have a shelf-life of 12 months

The chemical analysis instrument for electroplating baths

CDR GalvanLab®

  • Complete analysis panel, supplied already configured
  • Up to 16 determinations simultaneously
  • Possibility of carrying out analyses of the same sample
  • Integrated printer
  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)