Determination of Total Organic Carbon in process water
TOC analysis allows the quantification of Total Organic Carbon in galvanic process water. Monitoring this parameter in water is important to ensure the high quality standards of the electroplating industry required by the market.
Method
Test type: End Point.
Time testing: 10 minutes
Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab®.
Sample treatment
Use water as it is.
Test principle
Total Organic Carbon content is determined by its reaction with a chromogen that develops a coloration, the optical intensity of which is proportional to the concentration of Organic Carbon in the analyzed sample.
Calibration curve
The calibration curve was obtained by correlating the CDR method with the reference method.
Measuring range
| Analyses | Measuring range | Resolution | Repeatability |
|---|---|---|---|
| Total Organic Carbon [5-100] mg/L | 5 - 100 mg/L | 1.000 mg/L | 5 mg/L |
Reagent test Kits
Code *400140: 100 test
Code *400145: 10 test
The reagents have a shelf-life of 18 months
The chemical analysis instrument for electroplating baths
CDR GalvanLab®
- Complete analysis panel, supplied already configured
- Up to 16 determinations simultaneously
- Possibility of carrying out analyses of the same sample
- Integrated printer
- Full connections (LAN - USB - Bluetooth barcode/QR code reader)